Accuracy:
± 5 % of reading or ±0.035 mg/L (ppm), whichever is greater, from 0 - 6 mg/L (ppm)
Accuracy 2:
± 10 % of reading from 6.0 - 10 mg/L (ppm)
Alarm:
On screen display only
Cycle Time:
75 seconds per measurement
Dimensions (H x W x D):
380 mm x 336.2 mm x 203 mm
Display:
Graphic dot matrix LCD with LED backlighting, transreflective
Drain Fitting:
3/8 in. FNPT gravity drain
Enclosure Construction:
Powder coated steel
Enclosure Rating:
IP64
Flow Rate:
125 mL/min
Inlet Pressure:
4 - 6 psi
Manual Languages:
English, German
Mounting Configurations:
Wall Mount
Operating Humidity:
Max = 90 % at 50°C, non-condensing
Operating temperature range:
-5 - 50 °C
Outputs:
4 to 20 mA. 4-wire configuration with common-referenced current sink
Remote Interface:RS485 Modbus RTU
Power requirements (Voltage):
12 V DC
Range:
0 - 10 mg/L (ppm) as Chlorine† †The measured TRO concentration can be expressed as Bromine by choosing the option in the instrument menu.
Regulatory:
CE marked
Inclination: IEC-60092-504
Vibration: IEC-60068-2-6
ATEX version only: Ex II 2G Ex ib IIC T5 Gc (5°C ≤ Tamb ≤ 50°C)
Sample Flow Rate:
125 mL/min
Sample Inlet Connection:
1/4 in. FNPT sample line
Sample Pressure:
4 - 6 psi
Sample requirements:
Sample must be filtered to 35 micron or less
Sample Temperature:
5 - 40°C (41 - 104°F)
Storage conditions: