The Hach 1735 TRO Analyzer is an ATEX Zone 1 intrinsically safe certified analyzer for use in hazardous, or potentially hazardous environments. This design allows the user to interact and perform maintenance with ease. In addition, it does not require heavy enclosures or additional resources, such as purge gas.

Superior Reagent Life

Suitable for high temperature environments such as engine room installations on ships where the ambient temperature can rise as high as 50 °C. Hach reagents can provide accurate measurements for up to 70 days at these high temperatures. The extended reagent life minimizes interaction with the instrument and the operating costs.

Accurate Results

The Hach 1735 TRO Analyzer uses colorimetric DPD chemistry to monitor water online for TRO concentration. This is the same method as used in the Hach CL17 Chlorine Analyzer, an industry leader for 35 years. The accuracy of this analysis method is not affected by changes in sample pH, temperature, chlorine concentration, salinity (within the specified measurement ranges), pressure or flow, thus offering more accuracy than other available methods.

Minimal Maintenance

The Hach 1735 TRO Analyzer is designed to require minimum maintenance. Under typical use*, the Hach 1735 TRO Analyzer will operate unattended for 70 days. Less than 5 minutes of routine maintenance, including reagent changes, is required every 2 months. No calibration is needed, and no special tools are required. * 24 hours operation per week

Marine Application Ready

To measure total residual oxidant on vessels, the hach 1735 TRO Analyzer is built to withstand harsh shipboard conditions. It meetsIMO standards for tilt and vibration, and is protected by an IP64 rated enclosure which prevents corrosion caused by seawater.

Specifications

Accuracy: ± 5 % of reading or ±0.035 mg/L (ppm), whichever is greater, from 0 - 6 mg/L (ppm)
Accuracy 2: ± 10 % of reading from 6.0 - 10 mg/L (ppm)
Alarm: On screen display only
Cycle Time: 75 seconds per measurement
Dimensions (H x W x D): 380 mm x 336.2 mm x 203 mm
Display: Graphic dot matrix LCD with LED backlighting, transreflective
Drain Fitting: 3/8 in. FNPT gravity drain
Enclosure Construction: Powder coated steel
Enclosure Rating: IP64
Flow Rate: 125 mL/min
Inlet Pressure: 4 - 6 psi
Manual Languages: English, German
Mounting Configurations: Wall Mount
Operating Humidity: Max = 90 % at 50°C, non-condensing
Operating temperature range: -5 - 50 °C
Outputs: 4 to 20 mA. 4-wire configuration with common-referenced current sink
Remote Interface:RS485 Modbus RTU
Power requirements (Voltage): 12 V DC
Range: 0 - 10 mg/L (ppm) as Chlorine† †The measured TRO concentration can be expressed as Bromine by choosing the option in the instrument menu.
Regulatory: CE marked
Inclination: IEC-60092-504
Vibration: IEC-60068-2-6
ATEX version only: Ex II 2G Ex ib IIC T5 Gc (5°C ≤ Tamb ≤ 50°C)
Sample Flow Rate: 125 mL/min
Sample Inlet Connection: 1/4 in. FNPT sample line
Sample Pressure: 4 - 6 psi
Sample requirements: Sample must be filtered to 35 micron or less
Sample Temperature: 5 - 40°C (41 - 104°F)
Storage conditions: -20 °C to 60 °C