9187sc Chlorine Dioxide Analyzer with sc100 controller and RS232

OBSOLETE ITEM

9187sc Chlorine Dioxide Analyzer with sc100 controller and RS232
Product #: 6043402

OBSOLETE ITEM

Selective Membrane Avoids Interference from other species

The Hach 9187sc Amperometric Chlorine Dioxide sensor uses a membrane that is selective for chlorine dioxide species. There are no interferences from bromine, chlorine or hydrogen peroxide. The only potential interference is from ozone, which is rarely present in water treated with chlorine dioxide.

Flexible Options

Options available for the 9187sc ClO2 sensor include: Acidification Unit, which can be used intermittently or continuously for cleaning and is fully programmable, and Intermittent Flow Unit, which is fully programmable and saves resources while limiting excess flow to drain.

Easy Setup and Maintenance

The all-inclusive 9187sc ClO2 system comes pre-assembled on a panel. The membrane system of the sensor means there are no chemicals used for measurement. Maintenance is minimal and two years of typical maintenance items are included with the system.

Compatible with Hach Multi-Sensor, Multi-Parameter Digital Controllers

The Hach 9187sc ClO2 sensor can be used with Hach's new sc100 or sc1000 Digital Controllers. The sc100 controller accepts up to two sensors. The sc1000 accepts up to eight sensors. Multiple controllers can be networked to accommodate many more sensors and parameters, reducing the cost per measuring point. Just plug in any Hach "Plug & Play" digital sensor and it's ready to use without software configuration. "Plug & Play" connectivity means there's no complicated wiring or set up. Network the 9187sc ClO2 sensor with any of Hach's digital sensors for measuring dissolved oxygen, turbidity, ORP, conductivity, and many other parameters.

What's included?

9187sc Chlorine Dioxide Sensor, 9187sc User Manual, sc100 Controller with Modbus RS232, sc100 User Manual, sc100 Mounting Panel

Specifications

Accuracy: 5 % or ±10 ppb ClO2, whichever is greater
Calibration: Comparison of lab method with process sample
Calibration Interval: 2 months
Calibration Method: Electrically or with de-chlorinated water
Certifications: UL, CSA (certified by ETL), CE
Connection Drain Line: 1/2-in. I.D. (supplied)
Connections: 1/4-in. O.D.
Flow: 14 L/h (200 - 250 mL/min) au-regulated by flow thru cell
Interferences: Ozone
Lower Limit of Detection (LOD): Minimum 10 ppm or 0.01 mg/L ClO2
Maintenance Interval: Measurement Cell: 6 months for membrane and electrolyte, typical
Material: Electrodes: gold cathode/silver anode
Measuring Principle: Amperometric/Membrane (electrode,
membrane, electrolyte)
Mounting: Flat, vertical surface (panel, stand, etc.)
Operating temperature range: 0 - 45 °C
Pressure Range: 0.1 to 2 bar in flow cell
Range: 02 mg/L (mg/L) as chlorine dioxide (ClO2) 2 ppm
Relative Humidity: 5 - 90 % non-condensing
Response time: 90 % in less than 90 seconds
Sample Input: 0.25 " outer diameter
Sample Output: 0.5 " inner diameter
Sample Temperature: 2 - 45°C (35.6-113 F)
Ship Wt. (lbs): 6.5 kg(14.3 lbs.)
Standard deviation: 1.5 %
Temperature Compensation: Automatic over sample temperature range
  Automatic over sample temperature range
Weight: 6.5 kg
What's included?: 9187sc Chlorine Dioxide Sensor, 9187sc User Manual, sc100 Controller with Modbus RS232, sc100 User Manual, sc100 Mounting Panel