Simple, automatic operation

The system auto-calibrates, cleans, primes and displays alarms on the front panel. A menu-driven interface allows you to set the frequency of these functions.

Easy and fast grab sample capability

The grab sample capability of the APA 6000 allows a quick analysis of samples taken from alternative sample streams. It also allows analysis of a known standard as a quality assurance check. The grab sample history menu enables you to view the date, time and results of the last eight grab samples taken.

Rugged, lightweight enclosure

The strong, lightweight ABS case protects the instrument and resists corrosion. The enclosure is rated NEMA 4X (indoor)/IP66 and has a provision for air purge.

Includes 4 user-defined PID control loops

What's included?

Hach APA6000 Analyzer, a one-month supply of reagents, Tool kit, Maintenance kit, Installation kit, Installation and Maintenance Manual, and parameter specific Manual. (The power cord is ordered separately.)

Specifications

Alarm: 14 programmable
Data Storage: days of data in AquaTrend® Interface
Range: 0.02 - 2.0 mg/L NH₃ as N
Recorder Outputs: Two built-in 4 to 20 mA outputs; output span programmable over any portion of the analyzer's range
Relay output: 2 relays, SPDT, unpowered
Sample requirements: Optionally capable of analyzing up to two samples streams (requires sample sequencing kit)
Sample Temperature: 5 - 50 °C (41 - 122 °F)
What's included?: Hach APA6000 Analyzer, a one-month supply of reagents, Tool kit, Maintenance kit, Installation kit, Installation and Maintenance Manual, and parameter specific Manual. (The power cord is ordered separately.)