Chemistries                 1300 887 735

SOLITAX sc High Range Suspended Solids and Turbidity Analyzer for Insertion in Pipes

OBSOLETE ITEM

SOLITAX sc  High Range Suspended Solids and Turbidity Analyzer for Insertion in Pipes
Product #: 6940400

OBSOLETE ITEM

Self-cleaning device prevents erroneous values

Excellent correlation to laboratory analysis

Sensor life extended by full serviceability

Easy one-point calibration

Any two SOLITAX sc sensors can be installed using one Hach sc100 Controller

What's included?

Turbidity & Suspended Solids probe, controller, adapter, flange, user manual

Specifications

Accuracy: Turbidity: Defined according to ISO/WD 13530. Suspended Solids of reading or ±0.001 NTU, whichever is greater
Accuracy 2: Defined according to ISO/WD 13530.Turbidity Suspended Solids Less than 5 % of reading (depends on homogeneity of municipal activated sludge)
Cable Length: 10 m (optional extension cables available)
Calibration: Turbidity Suspended Solids:Formazin or StablCal® Standard
Based on gravimetric TSS analysis with a correction factor procedure
Calibration Method: Turbidity Suspended Solids:Formazin or StablCal® Standard
Based on gravimetric TSS analysis with a correction factor procedure
Certifications: CE certified to EN 61326-1, EN 61326/A1, EN 61326/A2, EN 61010-1
Construction - Sensor Body (RD 240/260): Wiper
Sensor body: Silicon
Stainless steel or PVC
Detection Limit: Turbidity

Suspended Solids 0.001 mg/L
Diameter Sensor: (diameter x length)
Insertion sensor: 60 mm x(2.4 x)
Flow: 3 m/s (9.8 ft./s) maximum
Length: (diameter x length) Insertion sensor: 315 mm
Lower Limit of Detection (LOD): Turbidity

Suspended Solids 0.001 NTU
Material: Optics Carrier and Sleeve: stainless steel 1.4571 or black PVC
Wiper Arm: stainless steel 1.4581; Wiper Blade: silicone (standard) Optional: Viton (LZX578); Wiper Shaft: stainless steel 1.4571
Threaded cable fitting: stainless steel 1.4305 or white PVC
Stainless steel can be susceptible to corrosion at chloride concentrations ≥ 500 mg/l.
Max. cable length Sensor-to-Controller: 100 m
Measuring Range Turbidity: Suspended Solids 0.001 - 4000 NTU
Mounting: Fixed to the rim of the tank
  Through sidewall of a pipeline using a ball valve; minimum pipe size (4 in.) in carbon or stainless steel
Operating temperature range: 0 - 40 °C
Parameter: obs
Pressure Range: Sample Pressure 6 bar (87 psi) maximum
Range: Turbidity
Suspended Solids 0.001 mg/L 50 g/L
Repeatability: Defined according to ISO/WD 13530. Turbidity Suspended Solids Less than 3 %of reading (depends on the homogeneity of municipal activated sludge)
Repeatability Turbidity: Defined according to ISO/WD 13530. Suspended Solids. Less than1 %of reading
Response time: Initial response in 1 second
Sample Temperature: 0 - 40°C (32 - 104°F)
Signal Average Time: 1 second
Turbidity: Units of Measure Turbidity
Suspended Solids user selectable—NTU, FNU, or TE/F
Units: Units of Measure
Turbidity
Suspended Solids: User selectable—g/L, mg/L, ppm, or % solids
Weight Sensor: Insertion stainless steel: 2.4 kg (5.3 lb.)
What's included?: Turbidity & Suspended Solids probe, controller, adapter, flange, user manual