Analogue Electrodes Input (High Impedance) :
1 for combined pH electrode with temperature probe
Application:
pH and alkalinity in waters (ISO9963-1)
Beaker Volumes Accepted:
150 mL
Burette Motor Resolution:
40000 steps
Calibration Interval:
for pHC elcectrode, days & hours
Detection Limit:
10 mg/L CaCO3
Display Type:
LCD 128 x 64 pixel , backlit
Embedded Titrating / Dosing Unit(s):
1 motor
Enclosure Construction:
ABS and enameled steel
End Point Titration:
2 EP, preset at pH 8.20 & 4.50
Internal Calibration Storage Capacity:
Last calibration
Internal Data Storage:
Last 55 results
Languages:
Spanish, English, French, Italian
Manual Languages:
English, Spanish, French, Italian, Danish, Dutch, Polish, Portuguese and Swedish
Non-Volatile Memory:
Yes
Operating Humidity:
< 80 %
Operating Interface:
Keypad, 7 membrane keys, splash-proof, PET
Operating Temperature:
-10 - 50 °C
Operating temperature range:
15 - 40 °C
Parameter:
Alkalinity
pH Direct Measurement:
Yes
pH Electrode calibration:
1 - 3 points
pH Resolution:
0.01
Power requirements (Voltage):
24 VAC, 1.25 A, 30 VA
Power Supply:
Benchtop Power Supply
Printer:
Yes (Optional)
Range:
10 - 2000 mg/L CaCO3
From 10 up to 2000 mg/L CaCO3
Sample Changer:
No
Sample Stand:
Yes, magnetic stirring
Sample Volume:
Adjustable
Specific Technology:
pH Titration
Titrant:
0.1N HCl or H₂SO₄
Titrant Calibration:
Yes
Units:
meq/L, mmol/L, ppm & mg/L CaCO3; dGH & °dH; gpg; °e & °Clark. °F
Weight:
approx. 4 kg
What's included?:
Titrator comes with 1x10ml PP syringe, magnetic stirrer, sensor holder, electrode 50 11 T with temperature sensor, 10x PP sample cups, stirring bars and mouse.