Accuracy:
±5 % or ±10 ppb ClO2, whichever is greater
Calibration:
Comparison of lab method with process sample
Calibration Interval:
2 months
Calibration Method:
Electrically or with de-chlorinated water
Certifications:
UL, CSA (certified by ETL), CE
Connection Drain Line:
1/2-in. I.D. (supplied)
Connections:
1/4-in. O.D.
Cycle Time:
90 % in T<90sec
Flow:
14 L/h (200 - 250 mL/min) au-regulated by flow thru cell
Interferences:
Ozone
Lower Limit of Detection (LOD):
Minimum 10 ppm or 0.01 mg/L ClO2
Maintenance Interval:
Measurement Cell: 6 months for membrane and electrolyte, typical
Material:
Electrodes: gold cathode/silver anode
Measuring Principle:
Amperometric/Membrane (electrode,
membrane, electrolyte)
Mounting:
Flat, vertical surface (panel, stand, etc.)
Operating temperature range:
0 - 45 °C
Pressure Range:
0.1 to 2 bar in flow cell
Range:
- (0.010 mg/L -2 mg/L)
02 mg/L (mg/L) as chlorine dioxide (ClO2) 2 ppm
Relative Humidity:
3 - 90 % non-condensing
Response time:
90 % in less than 90 seconds
Sample Input:
0.25 " outer diameter
Sample Output:
0.5 " inner diameter
Sample Temperature:
2 - 45°C (35.6-113 F)
Ship Wt. (lbs):
6.5 kg(14.3 lbs.)
Standard deviation:
1.5 %
Temperature Compensation:
Automatic over sample temperature range
Automatic over sample temperature range
Weight:
6.5 kg
What's included?:
9187sc Chlorine Dioxide Sensor, 9187sc User Manual, sc100 Controller, sc100 User Manual, sc100 Mounting Panel