Prices Online Include GST          1300 887 735

CLF10 sc Free Chlorine Sensor, sc100 Controller, and Stainless Steel Panel with Grab Sample Only

OBSOLETE ITEM

CLF10 sc Free Chlorine Sensor, sc100 Controller, and Stainless Steel Panel with Grab Sample Only
Product #: 2979400

OBSOLETE ITEM

Compatible with Hach's "Plug and Play" Digital Controllers

The CLF10 sc and CLT10 sc analyzers leverage Hach's exclusive self diagnostics to alert users when the process has changed or the instrument needs servicing. Diagnostic features include the Cal Watch algorithm for warning of pH and chlorine calibration deviation and a non-contacting flow sensor for notification of insufficient sample flow.

EPA compliant according to Method 334.0

Chlorine measurement with an amperometric analyzer, such as the CLF10 sc or CLT10 sc, does not require reagents, eliminating the need for routine reagent replacement and waste stream management.

Exclusive Self Diagnostics

The CLF10 sc and CLT10 sc analyzers allow for real-time control of disinfection processes by providing continuous readings that indicate when treatment conditions have changed.

Compatible with Hach's "Plug and Play" Digital Controllers

The CLF10 sc and CLT10 sc analyzers can be used with any Hach sc digital controller. Whether you're measuring turbidity or chlorine, you only need to learn one controller for all your water analysis measurement points. Hach sc controllers have no complicated wiring or setup procedures. Just plug in any Hach digital sensor and it's ready to use without software configuration.

EPA Compliant According to Method 334.0

In accordance with EPA Method 334.0, the CLF10 sc and CLT10 sc analyzers can be used for reporting chlorine residual measurements. Additionally, Hach has created a suite of laboratory products and methods to help with startup and quality control procedures required in Method 334.0. (See page 4 for a partial listing of accessories.)

What's included?

Hach sc100 Controller, Mounting Hardware, sc100 Basic User Manual, CL10sc Panel, 1 M digital extension cable, Panel Manual, Chlorine Sensor Manual.

Specifications

Accuracy: Free Chlorine: ±3 % of the reference test** (DPD) at constant pH less than 7.2 (±0.2 pH unit)
Accuracy 2: Free Chlorine: ±10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 3: Total Chlorine: ±10 % of the reference test** (DPD) at stable pH less than 8.5 (±0.5 pH unit from the pH at calibration)
Accuracy 4: Total Chlorine: ±20 % of the reference test** (DPD) at stable pH greater than 8.5
Cable Connection: 5 pin, M12 connector
Cable Length: 1 m (between gateways and sc-controller)
Calibration Method: (zero and slope) calibration
Calibration Method 2: (zero and slope) calibration
Certifications: CE / ETL, EMC
Connection Drain Line: pH Flow Cell Outlet: 1/2-inch ID
Connections: 1/4 inch OD
Controller Compatibility: sc controller platform
Dimensions: mm x 25 mm (7.68 in x 0.98 in)
  Sensor Only: 195
Interferences: Monochloramine, chlorine dioxide, ozone,
and chalk deposits
Length: Sensor: 195 mm
Lower Limit of Detection (LOD): 30 ppb (0.03 ppm)
Material: Corrosion-resistant materials, fully-submersible
(stainless steel, PVC, silicon rubber and polycarbonate)
Measurement method: Reagentless, electrochemical, three-electrode amperometric system
Mounting: Panel or Wall Mounting
Operating temperature range: 0 - 45 °C
pH Range: 4 - 9
Power requirements (Voltage): 12 V DC
Pressure Range: 0.5 bar, no pressure impulses and/or vibrations
Range: 0 - 20 ppm Cl₂
Response time: Free Chlorine: 140 s or less for 90% change (T90)
at a stable temperature and pH
Response Time: Total Chlorine:100 sor less for 90% change (T90)
at a stable temperature and pH
Sample Flow Rate: 30 - 50 L/h (7.9 - 13.2 gal/hour),
Optimal is 40 L/hour (10.5 gal/hour)
Sample Temperature: (compensated for fluctuations) 5 - 45°C(41 - 113°F)
Sampling Time: Continuous
Ship Wt. (lbs): Approximately9.1
Temperature Compensation: Internal temperature sensor
  Internal temperature sensor
Warranty: 1 Year
Weight: 5.5 kg Approximately 5.5 kg (12 lbs)
Panel and empty panel-mounted components only
What's included?: Hach sc100 Controller, Mounting Hardware, sc100 Basic User Manual, CL10sc Panel, 1 M digital extension cable, Panel Manual, Chlorine Sensor Manual.