Chemistries                 1300 887 735

EZ1032 Phosphate Analyser (Molybdate Blue Method) PO₄-P

EZ1032 Phosphate Analyser (Molybdate Blue Method) PO₄-P
Product #: EZ1032.XXXXXXXX
Unit Price Contact Hach

What's included?

EZ1032 Phosphate Analyser, Instruction Manual, 1 x Double Bit Door key, 1 x Mounting Brackets, 1 x empty 5 Reagent Container with Fittings (for Colour Solution), and 1 x empty 2.5L Glass Reagent Container with Fittings (for Reductor Solution)

Specifications

Alarm: 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Ambient Temperature: 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing)
Analog Outputs: Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Automatic cleaning: Yes
Calibration: Automatic, 2-point; frequency freely programmable
Certifications: CE compliant / ETL certified
Cycle Time: 10 min (dilution + 5 min)
Demineralised water: For rinsing / dilution
Digital outputs: Optional: Modbus (TCP/IP, RS485)
Dimensions (H x W x D): 690 mm x 465 mm x 330 mm
Drain: Atmospheric pressure, vented, min. Ø 32 mm
Earth connection: Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²
Instrument air: Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Interferences: Arsenic (V), Chromium (VI), Copper (II) > 10 mg/L, Iron (III) > 10 mg/L, Sulphide > 2 mg/L, Vanadium, Silica > 60 mg/L. Large amounts of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar.
Lower Limit of Detection (LOD): ≤ 1 µg/L
Material: Hinged part: Thermoform ABS, door: PMMA
Wall section: Galvanised steel, powder coated
Measurement method: Molybdate blue method (630 nm), conform with APHA 4500-P (C) and ( E)
Model: EZ1032
Number of sample streams: 1, 2, 4, or 8
Parameter: Orthophosphate (PO₄-P)
Power: 100 - 240 VAC, 50/60 Hz
Max. power consumption: 120 VA
Power supply: 100 - 240 VAC, 50/60 Hz
Precision: 1% full scale range for standard test
Protection Class: Protection class: Analyser cabinet: IP44 / Panel PC: IP65
Range: 0.005 - 1 mg/L PO₄-P
  0.005 - 1 mg/L PO₄-P

Optional:
0.001 - 0.1 mg/L
0.0025 - 0.25 mg/L
0.0025 - 0.5 mg/L
0.04 - 4 mg/L (with internal dilution)
0.08 - 8 mg/L (with internal dilution)
0.5 - 100 mg/L (with internal dilution)
Reagent Requirements: Keep between 10 - 30 °C
Sample Flow Rate: 100 - 300 mL/min
Sample Pressure: By external overflow vessel
Sample Quality: Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU
Sample Temperature: 5 - 30 °C
Validation: Automatic; frequency freely programmable
Warranty: 12 months
Weight: 25 kg
What's included?: EZ1032 Phosphate Analyser, Instruction Manual, 1 x Double Bit Door key, 1 x Mounting Brackets, 1 x empty 5 Reagent Container with Fittings (for Colour Solution), and 1 x empty 2.5L Glass Reagent Container with Fittings (for Reductor Solution)