Accuracy:
With Standard Solution:
Measuring range 1: ≤ 1 mg/L: 3% of the measured value + 0.02 mg/L > 1 mg/L: 5% of the measured value + 0.02 mg/L
Measuring range 2: 3% of the measured value + 0.05 mg/L
Measuring range 3: 3% of the measured value + 1.0 mg/L
Measuring range 4: 4.5% of the measured value + 10 mg/L
Altitude:
2000 m maximum
Cable Length:
Data and power cable: 2 m from edge of enclosure
Certifications:
CE, UKCA, CMIM, FCC, ISED, certified to UL and CSA safety standards by TÜV
Controller Compatibility:
Digital SC Controller. Controllers are sold separately.
Dimensions:
575 x 991 x 425 mm (22.63 x 39.01 x 16.73 in.)
Enclosure Rating:
Rating: IP55, NEMA 3R UL50E
Material: PUR 66
Filtersystem included:
Integrated sample supply pump for FX620 Filtration System
Indoor/Outdoor:
Indoor or outdoor use
Lower Limit of Detection (LOD):
Measuring range 1: 0.02 mg/L NH₄-N
Measuring range 2: 0.05 mg/L NH₄-N
Measuring range 3: 1 mg/L NH₄-N
Measuring range 4: 10 mg/L NH₄-N
Material:
PUR 66 UV-resistant
Measurement method:
Gas sensing electrode (GSE)
Measurement Range 2:
0.05 - 20 mg/L NH₄-N
Measurement Range 3:
1 - 100 mg/L NH₄-N
Measuring Interval:
5, 10, 15, 20 or 30 min. (user adjustable)
Measuring range:
10 - 1000 mg/L NH₄-N
Number of Channels:
1
Operating temperature range:
-20 - 45 °C; 95% relative humidity, non-condensing
Options:
Flow detection
Outputs:
Relay, analog outputs, network interface through SC Controller
Parameter:
Ammonium
Permissible Chloride Range:
≤ 1000 mg/L Cl-
pH Range:
5 - 9 pH
Pollution Degree:
2
Power supply:
Power supplied by the SC controller. With the SC4500, an additional power box is needed.
Analyser and heated drain tubing (powered by analyser): 115 VAC or 230 VAC
Protection Class:
Class 1
Range:
0.02 - 5.0 mg/L NH₄-N
Repeatability:
With Standard Solution:
Measuring range 1: 3% of the measured value + 0.02 mg/L
Measuring range 2: 2% of the measured value + 0.05 mg/L
Measuring range 3: 2% of the measured value + 1.0 mg/L
Measuring range 4: 2% of the measured value + 10 mg/L
Response Time T90:
Measuring range 1: ≤ 0.2 mg/L: 3 measurements required (15 total minutes minimum); > 0.2 mg/L: 1 measurement required (5 total minutes)
Measuring range 2: < 5 minutes
Measuring range 3: < 5 minutes
Measuring range 4: < 5 minutes
Sample Flow Rate:
1.0 - 20.0 L/h to the analyser from the external sample filtration
Sample Pressure:
2.5 bar (0.25 mPa) maximum
Sample Quality:
Ultra filtrated or comparable (FX6x0 < 0.45 µm)
Sample Temperature:
4 - 40 °C
Special Remarks:
The liquid level in the basin must be below the bottom of the analyser
Storage conditions:
-20 - 60 °C; 95% relative humidity, non-condensing
Weight:
Approximately 45 kg without chemicals