Calibration Interval:
1 year; calibration time 90 minutes
Calibration Method:
Automated work: 18-point calibration using KHP (6-point concentration, repeated 3 times)
Carrier Gas Options:
Air without CO², O², or N²
Compliance Certifications:
ISO 8245 and DIN EN 1483;
● United States Pharmacopeia (USP) (including Sterilized Water SST), European Pharmacopoeia (EP), Japanese Pharmacopoeia (JP), IP, CP, KP;
● Clean Water Test Method, US EPS5310c;
● FDA 21 CFR, part 11 - United States Pharmacopoeia (USP), European Pharmacopoeia (EP), Japanese Pharmacopoeia (JP), IP, CP, KP.
Data Export:
PDF, CSV
Dimensions:
410 mm x 320 mm x 507 mm
Display Type:
10.4" high resolution color touch screen
Inorganic Carbon Handling:
No external inorganic carbon removal module required
Overload Recovery:
1 measurement
Oxidation Method:
UV lamp + persulfate
Particle Size:
Up to 100 µm
Power requirements (Hz):
47 - 63 Hz
Power requirements (Voltage):
100 - 240 VAC
Precision:
3% or 3 ppb, whichever is greater
Range:
0.4 ppb - 100 ppm
Sample homogenisation:
An autosampler is also available
Sample to Sample Carryover:
< 0.2%
Warranty:
12 months