Accuracy:
5% of measured value + 0.2 mg/L (with standard solutions) NH4-N
Cable Length:
10 m / 32.8 fixed cable at sensor
10 m fixed cable at sensor, available up to 100 m
Calibration Method:
Sensorcode (calibrated sensor cartridge) or 2-point in-line value and matrix correction
Controller Compatibility:
sc 200, sc 1000
Data Transfer:
current output, relays and bus communication via sc controller
Flow:
< 4 m/s max.
Includes:
Ammonium probe, sensor cartridge, user manual
Installation Angle:
45 ° ± 15 ° (vertical in flow direction)
Length:
320 mm
Lower Limit of Detection (LOD):
0.2 mg/L NH4–N
Material:
Cartridge: Stainless steel (1.4571), PVC, POM, ABS, NBR
Measurement method:
Potentiometric ion-selective measurement
Measurement Range:
±5% of the measured value + 0.2 mg/L (with standard solution) NH4-N
Measuring Interval:
Continuous
Measuring uncertainty:
0.2 mg/l NH4-N (with standard solutions)
Operating temperature range:
Air: -20 - 45 °C (-4 - 113 °F)
Parameter:
NH4-N
pH Range:
5 - 9 pH
Process Connection:
1 inch thread
Protection Class:
IP68
Range:
0 - 1000 mg/L NH4-N
Repeatability:
±5 % of the measured value (with standard solution) NH4-N
Response time:
< 3 min
Sample Pressure:
0.3 bar max.
Sample Temperature:
2 - 40 °C (35.6 - 104 °F)
Sensor Cartridge:
With CARTRICAL plus technology: compact housing containing calibrated electrodes for ammonium, potassium, reference system and temperature sensor, all calibrated to each other; typical lifetime six months
Sensor Immersion Depth:
0.3 - 3 m (1.0 - 10 ft)
Storage conditions:
5 to 40 °C (41 to 104 °F) sensor cartridge
Warranty:
12 months
Weight:
2.38 kg
Wetted Materials:
Probe: Stainless Steel 1.4571; ASA + PC, silicon; PVC and PU;
Sensor cartridge: PVC; POM; ABS; Stainless Steel 1.4571; NBR
What's included?:
Ammonium probe, sensor cartridge, user manual