Beam Height:
10 mm
Data Logger:
2000 measured values (Result, Date, Time, Sample ID, User ID)
Dimensions (H x W x D):
151 mm x 350 mm x 255 mm
Display:
7" TFT
Display Resolution:
WVGA (800 pix x 480 pix)
Display Size:
7 inch (17.8 cm)
Display Type:
Colored touch-screen
Enclosure Rating:
IP40
Interfaces:
USB type A (2), USB type B, Ethernet, RFID module
Light source:
Gas-filled Tungsten (visible)
Manual Languages:
English
French (CDN)
Spanish (SA)
Portuguese (BR)
Chinese
Japanese
Korean
Operating Conditions:
10 - 40 °C (50 - 104 °F), max. 80 % relative humidity (non-condensing)
Operating Mode:
Transmittance (%), Absorbance and Concentration, Scanning
Optical System:
Reference beam, spectral
Photometric Accuracy:
5 mAbs @ 0.0 - 0.5 Abs
1 % at 0.50 - 2.0 Abs
Photometric Linearity:
< 0.5 % - 2 Abs
Photometric Linearity 2:
≤ 0.01 % at >2 Abs with neutral glass at 546 nm
Photometric Measuring Range:
± 3.0 Abs (wavelength range 340 - 900 nm)
Power Requirements:
With external power supply
Power requirements (Hz):
50/60 Hz
Power requirements (Voltage):
110 - 240 V AC
Power Supply:
Benchtop Power Supply
Preprogrammed Methods:
> 240
Sample Cell Compatibility:
13 mm round, 16 mm round, 1 cm & 5 cm rectangular, 1" round, 1" rectangular
Scanning Speed:
> 8 nm/S (in steps of 1 nm)
Source Lamp:
Gas-filled Tungsten (visible)
Specific Technology:
RFID for easy method update, sample ID and Certificate of Analysis
Spectral Bandwidth:
5 nm
Standard accessories:
None
Storage conditions:
-30 - 60 °C (-22 - 140 °F), max. 80% relative humidity (non-condensing)
Stray Light:
< 0.1% T at 340 nm with NaNO2
User Interface Languages:
Bulgarian, Chinese, Croatian, Czech, Danish,
Dutch, English, Finnish, French, German,
Greek, Hungarian, Italian, Japanese, Korean,
Polish, Portuguese - Brazil, Portuguese,
Russian, Serbian, Slovakian, Slowenian,
Spanish, Swedish, Turkish
User Programs:
100
Warranty:
12 months
Wavelength Accuracy:
± 1.5 nm (wavelength range 340 - 900 nm)
Wavelength Calibration:
Automatic
Wavelength Range:
320 - 1100 nm
Wavelength Reproducibility:
± 0.1 nm
Wavelength Resolution:
1 nm
Wavelength Selection:
Automatic, based on method selection
Weight:
4.2 kg
What's included?:
Includes: Adapter A for 1 in. round and 1 cm square cells, matched pair of 1 in. square glass sample cells, light shield, dust cover, printed basic user manual and benchtop power supply with 115 and 230V power cords.