Ambient Temperature:
5 - 45 °C
Channels:
2 Channel / Grab sample - Calibration port
Communication:
Modbus RTU, Modbus TCP/IP & Profibus (when the Profibus option is selected, the digital output signals are sent through the Profibus converter with its specific communication protocol)
Cycle Time:
From 5.5 minutes, depending on range and application
Data Storage:
Previous 9999 reaction data
Previous 99 fault events
Display:
High contrast 40 character x 16 line backlit LCD with LED backlight
Humidity:
5 - 85 % (non-condensing)
Languages:
English, Spanish, French, German
Others available upon request.
Measurement method:
Infrared measurement of CO₂ after oxidation
(DIN EN 1484:1997-08, ISO 8245:1999-03, EPA 415.1)
Multi-Stream:
2 stream
Number of streams:
Dual stream
Outputs:
Three 4-20 mA outputs enabled as standard
Oxidation Method:
Unique Two-Stage Advanced Oxidation Process (TSAO) using Hydroxyl Radicals, Hach Company method 10261 (EPA approved for drinking water)
Parameter:
TOC (NPOC), TIC, % TOC removal (with 2 streams) and COD, BOD after correlation
Particle Size:
Up to 100 µm
Power requirements (Hz):
50 Hz
Power requirements (Voltage):
230 V AC
Range:
0 - 25 mg/L C
Repeatability:
±3% of reading or ±0.03 mg/L, whichever is greater;
Lower Limit of detection LOD = 0.06 mg/L
Sample Inlet Temperature:
0 - 60 °C
Service Interval:
6 months service intervals
User Interface:
Microcontroller with membrane keyboard
Warranty:
12 months
Weight:
46 kg
What's included?:
Includes: B3500dw analyser, Tubing, Fuses, Ferrules, Drain, Acid & Base Dip Tubes, C02 Filter & B3500dw User Manual