Accuracy:
± 0.05 ppm Below 5 ppm
Accuracy 2:
± 0.1 ppm Above 5 ppm
Accuracy 3:
Temperature: ± 0.2 °C
Body material:
Stainless
Cable Length:
10 m (33 ft)
Calibration Method:
All Probes are factory calibrated and ready to use
Air Calibration: one point, 100% water saturated air
Sample Calibration: comparison to standard instrument, or comparison to Winkler Titration method
Communication:
Modbus
Controller Compatibility:
SC4500, SC1000
Diameter:
48.25 mm
Dimensions (D x L):
48.25 mm x 254 mm
Flow Rate:
None required
Interferences:
No Interferences from the following: H2S, pH, K+, Na+, Mg2+, Ca2+, NH4++, Al3+, Pb2+, Cd2+, Zn2+, Cr (total), Fe2+, Fe3+, Mn2+, Cu2+, Ni2+, Co2+, CN-, NO3-, SO42-, S2-, PO43-, Cl-, anion active surfactants, crude oils, Cl2 < 4 ppm
Length:
254 mm
Measurement Range 2:
0 - 20.00 mg/L
Measurement Range 3:
0 - 200 %
Options:
Mounting Adapter
Range:
0 - 20.00 ppm
Repeatability:
± 0.1 ppm (mg/L)
Resolution:
0.01 ppm (mg/L)
0.1% saturation
Response time:
T90 < 40 s
T95 < 60 s
Sensor Immersion Depth:
Down to 34 m (112 ft.), 345 kPa (50 psi), maximum; accuracy may not be maintained at this depth
Storage conditions:
-20 °C to 70 °C
Transmission Distance:
1000 m (3280 ft.) maximum when used with a termination box
100 m (328 ft.) maximum
Warranty:
36 months
Weight:
1 kg (2.2 lbs), probe only
What's included?:
Includes: LDO probe, mounting adapter, and manual