Chemistries                 1300 887 735

NITRATAX plus sc 2MM W/sc200

OBSOLETE ITEM

NITRATAX plus sc 2MM W/sc200
Product #: 2984000

OBSOLETE ITEM

UV Analysis Eliminates Reagents, Sampling, and Sample Conditioning

Using advanced ultraviolet (UV) absorption technology, Hach NITRATAX sc UV Nitrate Sensors offer unprecedented simplicity, accuracy, and economy in nitrate analysis. By continuously measuring the UV light absorbed by nitrates, NITRATAX sc sensors eliminate the need for reagents, sample conditioning, and frequent calibrations.

UV Absorption Method - Proven, Continuous, Precise

NITRATAX sc sensors rely on the principle that molecular bonds absorb ultraviolet (UV) light. In this case, nitrate (NO3) and nitrite (NO2) absorb UV light. As the concentration of nitrate or nitrite increases, UV absorption also increases. A built-in photometer measures the primary beam, while a second beam of UV light provides a reference standard and corrects for interference caused by turbidity and organic matter. Results are reported on the controller unit's alphanumeric display in mg/L of nitrate (NO3) or nitrate nitrogen (NO3-N).

Choice of Three Models

NITRATAX plus sc Process sensor for continuous measurement in drinking water, wastewater, or activated sludge. Turbidity compensation using reference measurement. NITRATAX eco sc low cost sensor for measurement, especially for sewage treatment plants with intermittent aeration technology. Turbidity compensation using reference measurement. NITRATAX clear sc Process sensor for continuous measurement in clean water such as drinking water or wastewater effluent.

Self-Cleaning Sensor

With the sensor submerged in the sample stream, the detector windows are automatically cleaned by a built-in wiper that eliminate surface films or particles that can diminish accuracy.

Optional Bypass Panel

An optional Bypass Panel or flow-through sample cell is available for clean water applications when direct immersion in Optional Bypass Panel a sample stream is impractical. The Bypass Panel uses the same probes as the immersion sensors for greater flexibility.

What's included?

Nitrate probe, controller, installation kit, standard solution, user manual

Specifications

Accuracy: 3 % of measured value +0.5 mg/l (with standard solutions)
Ambient Temperature: 2 - 40 °C
Cable Length: 10 m fixed cable at sensor
Diameter: 70 mm
Length: 333 mm
Lower Limit of Detection (LOD): 1 mg/L to 100
Measurement method: Patented 2-beam method
Measuring Interval: 1 to 30 min (fixed values selectable)
Measuring Principle: UV absorption measurement
Operating temperature range: 2 - 40 °C
Power Consumption: 2 W
Pressure Range: Sensor pressure limit, 0.5 bar
Range: 0.1 - 50 mg/L NO2+3 -N
Response time: T 100, 1 min
Sludge Compensation: Yes
Warranty: 12 months
Weight: 3.6 kg
What's included?: Nitrate probe, controller, installation kit, standard solution, user manual