Accuracy:
                                            
                                            
                                        
									    FNU: ±2% of reading plus 0.01 FNU from 0 - 1000 FNU
FAU: ±10% of reading from 20 - 10000 NTU
NTU: ±2% of reading plus 0.01 NTU from 0 - 1000 NTU, ±5% of reading from 1000 - 4000 NTU, ±10% of reading from 4000 - 10000 NTU
                                    
                                
    						
								
                                    
                                        
                                            
                                                Accuracy 3:
                                            
                                            
                                        
									    Absorbance: ±0.005 Abs from 0 - 1 Abs at 860 nm
Transmittance: 0.12% T from 10 - 100% T at 860 nm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Air purge:
                                            
                                            
                                        
									    Dry nitrogen or instrument grade air (ANSI MC 11.1, 1975)
0.05 L/s at 69 kPa (10 psig); 138 kPa (20 psig) max
Hose barb connection for 1/8-inch tubing
                                    
                                
    						
								
                                    
                                        
                                            
                                                Certifications:
                                            
                                            
                                        
									    CE, KC, RCM
                                    
                                
    						
								
                                    
                                        
                                            
                                                Communication:
                                            
                                            
                                        
									    USB
                                    
                                
    						
								
                                    
                                        
                                            
                                                Compliance:
                                            
                                            
                                        
									    ISO
                                    
                                
    						
								
                                    
                                        
                                            
                                                Data logging:
                                            
                                            
                                        
									    2000 total logs, includes reading log, verification log and calibration log
                                    
                                
    						
								
                                    
                                        
                                            
                                                Dimensions (H x W x D):
                                            
                                            
                                        
									    153 mm x 395 mm x 305 mm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Display:
                                            
                                            
                                        
									    17.8 mm colour touch screen
                                    
                                
    						
								
                                    
                                        
                                            
                                                Interface:
                                            
                                            
                                        
									    2 USB-A ports for USB flash drive, external printer, keyboard and barcode scanner
                                    
                                
    						
								
                                    
                                        
                                            
                                                Light source:
                                            
                                            
                                        
									    Light-emitting diode (LED) at 860 ± 30 nm
                                    
                                
    						
								
                                    
                                        
                                            
                                                Measurement method:
                                            
                                            
                                        
									    Nephelometric
                                    
                                
    						
								
                                    
                                        
                                            
                                                Model:
                                            
                                            
                                        
									    TL2360 ISO
                                    
                                
    						
								
                                    
                                        
                                            
                                                Operating temperature range:
                                            
                                            
                                        
									    0 - 40 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Power requirements (Amps):
                                            
                                            
                                        
									    3.4 A
                                    
                                
    						
								
                                    
                                        
                                            
                                                Power requirements (Hz):
                                            
                                            
                                        
									    50/60 Hz
                                    
                                
    						
								
                                    
                                        
                                            
                                                Power requirements (Voltage):
                                            
                                            
                                        
									    100 - 240 VAC
                                    
                                
    						
								
                                    
                                        
                                            
                                                Range:
                                            
                                            
                                        
									    0 - 10000 NTU (ISO)
                                    
                                
                                    
                                        
                                            
                                            
                                                 
                                            
                                        
									    FNU (Ratio on): 0 - 1000
FNU (Ratio off): 0 - 40
FAU (auto range): 20 - 10000
NTU (Ratio on): 0 - 10000 auto decimal
NTU (Ratio off): 0 - 40
EBC (Ratio on): 0 - 2450 auto decimal
EBC (Ratio off): 0 - 9.8
Absorbance (auto range): 0 - 2.00
Transmittance (%): 1.0 - 100
Degree (mg/L): 0 - 100
                                    
                                
    						
								
                                    
                                        
                                            
                                                Reading modes:
                                            
                                            
                                        
									    Manual or auto range, signal averaging on and adjustable or off, ratio on or off
                                    
                                
    						
								
                                    
                                        
                                            
                                                Regulatory:
                                            
                                            
                                        
									    Meets ISO 7027, DIN EN 27027, DIN 38404 and NFT 9033
                                    
                                
    						
								
                                    
                                        
                                            
                                                Repeatability:
                                            
                                            
                                        
									    ±1% of reading or 0.01 FNU/NTU, whichever is greater (under reference conditions)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Response time:
                                            
                                            
                                        
									    Signal averaging off: 6.8 seconds / Signal averaging on: 14 seconds (when 10 measurements are used to calculate the average)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Sample cell compatibility:
                                            
                                            
                                        
									    Round cells 95 x 25 mm (3.74 x 1 in.) borosilicate glass with rubber-lined screw caps
Note: Smaller sample cells (less than 25 mm) can be used when a cell adapter is used.
                                    
                                
    						
								
                                    
                                        
                                            
                                                Sample requirements:
                                            
                                            
                                        
									    25 mm sample cell: 20 mL minimum
0 to 70 °C (32 to 158 °F)
                                    
                                
    						
								
                                    
                                        
                                            
                                                Source Lamp:
                                            
                                            
                                        
									    LED
                                    
                                
    						
								
                                    
                                        
                                            
                                                Stabilisation time:
                                            
                                            
                                        
									    Immediately
                                    
                                
    						
								
                                    
                                        
                                            
                                                Storage conditions:
                                            
                                            
                                        
									    -20 to 60 °C
                                    
                                
    						
								
                                    
                                        
                                            
                                                Units:
                                            
                                            
                                        
									    FNU, FAU, NTU, EBC, Abs (absorbance), %T (% transmittance) and mg/L
                                    
                                
    						
								
                                    
                                        
                                            
                                                Warranty:
                                            
                                            
                                        
									    12 months
                                    
                                
    						
								
                                    
                                        
                                            
                                                Weight:
                                            
                                            
                                        
									    3.0 kg
                                    
                                
    						
								
                                    
                                        
                                            
                                                What's included?:
                                            
                                            
                                        
									    TL2360 Turbidimeter, silicone oil, oiling cloth, 1-inch sample cells (30 mL) with caps (6x), Gelex secondary turbidity standardization kit, Stablcal calibration kit, power supply, power cord, dust cover