Would you like to use our Product Configurator tool to configure this product before adding it to your cart? If not, you may add it directly to your cart.
Product #:2976600
OBSOLETE ITEM
Continuous, Automatic Early Warning Systems
Use the Hach UVAS plus sc UV Absorbance/Transmittance Sensor to continuously protect plant treatment processes from high influent organic loads. Operators can use the continuous readings of UV absorbance or transmission to watch for sudden changes in organic load that would require alternate treatment procedures.
Control Activated Sludge Processes
Activated sludge processes require precise balancing of organic load, aeration, and nutrients. Continuous trending of the organics in the system with the UVAS plus sc sensor can help operators know how to balance other factors resulting in cost and time savings.
Self-cleaning Wiper System
With the UVAS plus sc sensor submerged in the sample stream, the detector windows are automatically cleaned by a built-in wiper that eliminates surface films or particles that can diminish accuracy.
Monitor Efficiency of UV Disinfection Process
UV light transmittance (UVT) is critical in the delivery of dose in a UV reactor. The delivered dose is determined by, among other things, the UVT of the source water, the intensity of the UV lamps, and the flow rate of the water source. UVT can be affected by many factors, from a simple change in the seasons to storm events. Potential changes in UVT should be considered in a UV disinfection system for optimized dose delivery. Hach's UVAS plus sc is designed to provide continuous UVT measurement of pre-disinfected source water. Operational costs related to sampling for UVT may be reduced with continuous online measurement. Data can immediately be incorporated into the operation in real time.
Self Diagnostics and Easy Maintenance
Diagnostic routines built into the UVAS plus sc sensor reduce the need for extensive calibration and maintenance. Only semi-yearly inspection and replacement of the wiper and seals is needed.
What's included?
SAC probe, controller, user manual
Specifications
Application:Outlet; Drinking water; Surface water
Cable Length:10 m
Configurations:Sensor with SC200 Controller
Controller:SC200
Controller Compatibility:SC4500, SC1000, SC200
Diameter:70 mm
Length:333 mm
Maintenance Requirements:1 h/month in standard applications